Abstract
The ramifications of patient migration toward the foot of the bed in intensive care units are not well understood. Migration may cause shear and friction between the patient and the mattress, reduce elevation of the patient's torso, and require frequent repositioning of the patient. This study assesses how bed design impacts both the amount of migration that patients undergo during head section articulation to 30° and 45° and the extent of torso compression following the articulation.
Keywords:
hospital bed design, low back injuries, patient migration, patient repositioning, pressure ulcers, safe patient handling
THE impact of patient migration toward the foot of the bed in hospitalized patients is not well understood. There are at least 2 causes of this migration: gravity over time and bed design. During head-of-bed elevation, some bed designs may mechanically push patients toward the foot section of the bed. Whether caused by gravity over time or from bed design, migration may be deleterious to the patient due to increased shear and friction between the patient and the mattress and by a reduction in therapeutic torso angle. Patient migration may also be deleterious to the caregiver by requiring more pull-up in bed maneuvers, which have been shown to cause musculoskeletal disorders in health care professionals.
The prevention of pressure ulcers (PUs) in the health care environment has been a quality care focus for many years. There are intrinsic patient factors that contribute to skin breakdown such as age-related deterioration of skin, muscle, and nerve cells, conditions that affect perfusion such as vascular disease and diabetes, and circulation issues that reduce oxygenation levels in the skin and muscle.1–3 Extrinsic support surface–related PU risk factors, which include increased pressure, shear, friction, heat, and moisture, have also been identified.4–7 Patient migration that causes movement across the mattress has a high probability of increasing shear and friction forces between the patient's body and the support surface.1,8
Reducing shear and friction has long been a goal of nursing care with current clinical interventions including the use of sliding sheets, a “lift” rather than “slide” repositioning approach, and others.4–7 However, there is limited information regarding the role that bed design may have in reducing or increasing the sliding that occurs both during head-of-bed elevation and flattening of the bed.
Therapeutic head-of-bed elevation has been shown to reduce the effort to breathe during ventilator weaning,9 increase Pao2/Fio2 ratios,10 increase end-expiratory lung volume,10 and prevent aspiration of gastric contents into the lungs11 and is one of the first interventions in early mobilization protocols.12 When patients migrate toward the foot of the bed, even if the bed remains at the desired 45° position, the patient's torso may flatten to the extent that it no longer receives the benefits of the head-of-bed elevation. What is not currently understood is the effect that raising the head of the bed has on patient movement and torso compression.
Patient migration is problematic not just for the patient but also for caregivers. When a patient slides down the bed, a nurse must reposition the patient toward the head of the bed. Repositioning patients can be a repetitive task, with reported frequencies for a nurse as much as 10 times per shift13 or more than 20 times per week.14 The US Occupational Safety and Health Administration15 reports that, in 2010, nursing aides, orderlies, and attendants had the highest rates of musculoskeletal disorders with an incidence rate of 249 per 10 000 workers (total of 27 020 cases). This was more than 7 times the average across all industries. “Lifting a patient in bed” was a patient-handling task that was identified as “high risk” in its report. The American Nursing Association has recently developed standards for safe patient handling and mobility aimed at eliminating manual patient handling to create safer work environments for its members.16 Technology that reduces the need for “pull-up in bed” maneuvers, therefore, may be beneficial to both patients and caregivers.
Understanding patient migration may encourage better bed design and provide objective data to allow health care providers to make informed decisions about bed purchases. Reducing migration may reduce the incidence of nosocomial PUs, allow patients to maintain therapeutic head-of-bed positioning, and protect caregivers by reducing the number of times they have to reposition the patient. The objective of this study was to evaluate patient migration of nonobese individuals during head-of-bed articulations in 4 commonly used intensive care unit (ICU) beds with different head-of-bed pivot designs.
METHODS
Study approach
This study investigated how patient migration and torso compression were impacted by bed design as the head of bed was elevated and returned to the flat position, using motion capture methods recently developed by Kotowski and associates.17 This method used a motion capture system (MotionAnalysis, Santa Rosa, California) that quantified the instantaneous 3-dimensional position of 7 markers placed on the subject's body at specified bony anatomic locations and 6 markers on the bed (Figure). Video data were collected and analyzed using Cortex software (Motion Analysis, Santa Rosa, CA). During head-of-bed articulation, the anatomic markers were compared with the stable markers placed on each of 4 ICU beds to determine relative movement. Two different bed articulations were studied: (1) 0-30-0 head of bed raised from flat to 30° to flat and (2) 0-45-0 head of bed raised from flat to 45° to flat. Each articulation condition was repeated 5 times in a random order. In total, 40 trials were completed by each participant (4 bed types × 2 articulation types × 5 repeats).
Open in a separate window
Subjects
To mimic the hospitalized population, 12 healthy volunteers (6 males and 6 females) were recruited who met height and weight stratification requirements. The 6 height-weight classifications were as follows: (1) tall-average weight; (2) tall-overweight; (3) average height-average weight; (4) average height-overweight; (5) short-average weight; and (6) short-overweight. The height categories were defined by short—below 25th percentile (males 171 cm and 157 cm and 180 cm, females >166 cm) for males and females, respectively. Weight categories were defined on the basis of body mass index with average weight (
Procedure
Subjects received a full explanation of the study procedures that included instruction that they should try to act like a comatose patient during bed articulations and not to make any attempts to reposition themselves. The study was approved by the university institutional review board. The sequence of the beds was counterbalanced in random order for each subject, whereas the individual articulation conditions were randomized within each of the beds. Prior to each trial, subjects were anatomically aligned on each bed by making sure that the trochanter was positioned on the bed as designated by the manufacturer.
Outcome variables
The independent variables were bed types and articulation. Four different bed types were evaluated: bed A (Hill-Rom Progressa with StayInPlace Technology and integrated Therapy mattress) was designed with a head pivot that slides backward simultaneously with a head section that extends when the head of bed is raised; bed B (Hill-Rom Progressa without the StayInPlace feature with an integrated Therapy mattress) had the head pivot sliding with no lengthening of the head section when raising the head of bed; bed C (Linet Multicare with Virtuoso mattress, EU version) had a simple stationary head pivot where the head-of-bed section pivots around when the head of bed is raised; and bed D (Stryker InTouch with XPRT mattress) had 2 hinge points for the head pivot where the head-of-bed pivot changes from one pivot point to a second pivot at 20°.
The dependent variables consisted of the following: (1) net displacement quantified both how much movement toward the foot of bed occurred at the highest point during the articulation (net displacement—head-of-bed up) and the displacement at the end of the articulation when the bed returned to flat (net displacement—returned to flat); (2) cumulative movement was the total amount of movement of the anatomic site during the entire articulation (summation of any movement, down and up, during elevation of head of bed and when returning to flat); (3) torso compression was the maximum decrease in distance between the shoulder and trochanter markers during the articulation. Larger values indicated that torso compression was increasing. This measure provided an objective assessment of the amount of scrunching or balling up of the participant from head-of-bed elevation; (4) perceived sliding assessed the subject's perception of sliding during the articulation conditions as assessed at the end of the trial condition on an integral scale with 1 (none) to 10 (sliding several inches); (5) perceived discomfort was assessed at the end of the trial condition for any region in the body with an integral scale of 0 (no discomfort) to 10 (most unbearable pain imaginable); and (7) subject preference ranking allowed subjects to rank the 4 beds in order of best (or most desirable) to worst (least desirable) (from 1 to 4) after they had completed all 4 bed trials.
For each trial, the exact position of the trochanter and ankle was calculated by using the motion capture markers in the recorded data relative to the head, seat (thigh), and foot sections of the bed, respectively.
Statistical analyses
Descriptive statistics (means and standard deviations) were computed for all of the outcome variables as function of bed and bed by articulation interaction. A repeated-measures analysis of variance was completed on each dependent variable to identify whether there were significant effects of bed type and articulation type. Post hoc analyses in the form of Tukey standardized honest difference tests were conducted to identify the source of differences between beds.
RESULTS
Subject demographics
The 6 female subjects had an average age of 35.0 (SD = 11.5) years, standing height of 164.9 (SD = 5.9) cm, and average body weight of 63.1 (SD = 14.1) kg. The 6 male subjects were 23.8 (SD = 9.7) years old, 175.8 (SD = 9.5) cm tall, and weighed 77.3 (SD = 11.2) kg.
Study design measures
The 4 beds differed significantly in the amount of net displacement, cumulative movement, and torso compression (Table ).
Table 1.
Open in a separate window
Net displacement—Head-of-bed up
When the head of bed was fully raised during the head-of-bed articulation at 0-30-0, the trochanter migrated significantly more (P
Net displacement—Returned to flat
When the bed returned to flat, less net migration was seen than the head-of-bed up data. Mean trochanter distances on each of the 4 beds ranged from 1.1 to 2.3 cm during the 45° elevation, which was explained by patients moving back toward the head of the bed as the bed returned to flat.
Cumulative movement
Cumulative movement, which measured the entire movement of the anatomic marker during the articulation, was significantly higher (P
Torso compression
Bed A demonstrated significantly less torso compression than all other beds for both the 0-30-0 and the 0-45-0 articulations at 3.7 cm (P
Perceptions of sliding and discomfort
Beds A and B were perceived by subjects to have significantly less sliding than beds C and D (by about 50% relative amount). Discomfort was more variable. Although it is important to note that the response level at most indicated only mild pain. These values also reflect a few participants having no discomfort in any trial.
Subject preference rankings
When ranking the beds for preference, 9 of 12 subjects ranked bed A as most preferred (Table ). The least desirable bed was bed D, with 10 of 12 participants ranking bed D the worst. It is interesting to note that preference was similar to objective data outcomes.
Table 2.
Open in a separate window
DISCUSSION
Bed design, specifically head-of-bed pivot design, influenced how subjects migrated down the bed in this study. When subjects were articulated from flat to 45°, beds C and D demonstrated around 10 cm of migration at the peak of the articulation and subjects moved a total of 18 to 20 cm (cumulative migration) as the head of bed returned to flat. Bed A, which has a more complex head-of-bed pivot that elongates, had 86% less migration at the peak of articulation and 60% less cumulative movement than bed D, which, given the magnitude of the distance traveled, is likely a clinically significant finding.
Motion capture methods, which were developed by Kotowski and associates,17 provide objective video data that were analyzed to quantify movement of anatomic locations during head-of-bed articulation. Although actual patient data in a clinical setting may have been desired, actual patients would not have been able to undergo the 40 head-of-bed articulations and 4 bed transfers that were required for this testing. In addition, these methods, which were strengthened by the within subjects design, repeated conditions, and quantitative data, require a controlled laboratory environment.
This objective assessment provides the foundation to accurately assess the movement of subjects with the “best” bed minimizing migration, cumulative movement, and torso compression. Bed A was identified as the bed that minimized the most of the metrics, whereas bed D had the highest amount of migration at the head-of-bed up position, cumulative movement, and torso compression.
Bed A had 3 key characteristics of the frame design that minimized the outcome variables: (1) a complex head section pivot point; (2) a head section that lengthens as the head of bed rises; (3) knee auto contour (which is an automatic rising of the knee section during head-of-bed elevation). The complex pivot point and lengthening head section cradled the patient and accommodated the natural elongation of the torso during the articulation. The knee pivot auto contour also reduced migration of the patient toward the foot of the bed by providing a small “stop” for the thighs. The combination of these features was likely why bed A minimized both migration and torso compression. Bed B, which had a sliding head pivot but lacked lengthening of the head section, had significantly improved migration and cumulative movement as compared with beds C and D and was slightly worse that bed A.
As the head pivot design becomes simpler, the patient migration significantly increased during the articulation, which resulted in greater cumulative movement. Beds C and D had 4 to 5 times as much migration down the bed compared with beds A and B when the bed was articulated up to the upright position. This translated into 2 to 4 times as much cumulative movement. Bed C had a simple stationary head pivot, whereas bed D had 2 hinge points for the head pivot where head of bed pivot changes from one pivot point to a second pivot at 20°. In both cases, the head pivot articulation was simple and resulted in more movement of the patient (eg, pushed the patient down the bed). Both of these beds did not have any auto contour for the knee pivot during these traditional head-of-bed elevations. Even without auto contour, the articulations for beds C and D produced the most torso compression. This basically means that the shoulders are pushed toward the foot of the bed as well as down toward the lower torso and at a rate faster than the lower torso, compressing the torso by 7 to 8 cm.
Although shear and friction were not directly measured in this study, it is logical that a repetitive 18- to 20-cm movement across the mattress may increase the risk of PUs over time. The movement was measured at the trochanter and the ankle, which anatomically would demonstrate the movement for the buttocks and sacral regions and the heel, which are the most common areas for nosocomial PUs to develop.18
One of the benefits of reducing patient migration during head-of-bed elevation is the potential to limit manual lifting and patient-handling tasks, which are known causes of injuries and musculoskeletal disorders in nurses and other health care workers. The injuries sustained by health care workers have been reported as often life-altering and career-ending.16 Epidemiologic studies have found that repositioning patients accounts for between 25% and 50% of all low back injuries for nurses18–22 and has been identified as one of the worst tasks that nurses complete during their shifts.13,23–29 Biomechanical laboratory studies that estimated the spine loads on caregivers using sophisticated models have indicated that repositioning was among the worst patient-handling tasks performed by health care providers.30–33 All of these load estimates exceeded the recommended load tolerance estimates,34,35 indicating that the expected loads would be risky for the majority of the individuals performing these tasks (eg, health care providers).
A simple yet practical method to determine how much migration occurs when raising the head of a bed is to put a piece of tape on the bed that is lined up with the ankle, then raise the head of bed and measure how far down the ankle slid down the bed. A simple method for measuring torso compression would be to measure the original distance between the shoulder and the hip when the bed is in a supine position with a string and elevate the head of the bed with the string in place. Greater slack in the string would correspond to greater torso compression. With the importance of reducing PUs and low back injuries in caregivers, these simple metrics may lead to better outcomes and objective decision making.
CONCLUSION
Bed design significantly impacts the amount of migration during head-of-bed articulations. Migration, cumulative movement, and torso compression were higher on beds with simple head section pivot designs that lack the ability to elongate and slide backward. Increased migration may increase the risk of PUs for patients and increase the risk of low back injuries for caregivers. Torso compression may impact patient comfort. Designers of beds can use these results to improve on traditional and simple head-of-bed pivots.
A practical application of this study may be to visually assess the movement of the heels (to estimate migration) and assess the change in distance between the shoulders and the trochanter (to estimate torso compression) when deciding to purchase ICU beds. However, objective measurements are the only true way to understand the impact of head section articulation on a patient.
Footnotes
This project was funded by Hill-Rom, Inc.
The authors declare there is no conflict of interest. The funding was received through normal research channels, and neither author has a vested interest in Hill-Rom, Inc.
What do I need for a Dolby Atmos setup?
The first thing you'll need for a 5.1.2 Dolby Atmos home theater setup is a Dolby Atmos capable receiver. Additionally, you'll need eight speakers: a center speaker, two left and right main speakers, left and right surrounds, a pair of Dolby Atmos-enabled speakers, and a subwoofer.
You'll also need to handle the visuals, whether that be from an external device like a blue-ray player, a projector, or the internal apps on your TV.
Lastly, you'll need the proper cables. High-speed HDMI cables, enough speaker wire to connect all of the speakers, and an RCA cable long enough to connect your subwoofer to your receiver.
We know this seems like a lot, but don't worry. For receivers, you can consult our AV Receiver Buying Guide and gear guide to help you decide on the best option. If you're looking for a new TV to compliment your new Dolby Atmos system, we've got your back there, too! Check out our guide on how to choose which TV to buy to see all our suggestions on how you can pick the TV that meets your needs no matter what your budget is.
How much does a home theater cost?
It really depends on what you want, the size of the room, and your budget. A home theater setup can vary from $667 to $8,000 or more. Fortunately, we have home theater bundles for every price range!
What home theater receiver do I need?
That depends on what the rest of your setup is like. We'd recommend taking a look at our AV receivers Buying Guide to find out what to look for to fit your needs! If you're still having trouble, you can always reach out to us, with over a dozen seasoned pros standing by to help you figure out precisely what you need!
What home theater system should I buy?
You should buy what fits your wants and needs, and never settle for less! For a good starting point, you can take a look at some of our recommendations here.
What home theater projector should I buy?
We'd suggest you take a look at our guide for Best Projectors. There are plenty of options there for different needs and budgets; we're confident you'll find one that fits you. If you're still feeling a little lost, you can always contact us and one of our pros will be able to give you a recommendation based on your needs.
Can I connect my home theater to my laptop or PC?
It depends on the system, but if it can be set up through a single HDMI cable or wirelessly, it can be!
Are home theater systems good for music?
Heck yeah they are! While music might not be as immersive as movies or TV shows with a good surround system, it can still certainly provide a fully immersive surround listening experience.
Are Home Theater Systems Wireless?
Not all of them, but there are some wireless options available!